Find the non-invasive respiratory support you need to help patients come home from the hospital, with Vivo 30, featuring Pressure Control and Pressure Support modes.


Features:

Bring advanced and clinically reliable respiratory solutions to patients in the hospital or at home with Vivo 30, a Bi-Level ventilator offering non-invasive respiratory support through Pressure Control and Pressure Support modes.

Vivo 30 offers advantages for both patients and providers:

eSync, highly responsive Trigger technology, enables Vivo 30 to synchronize smoothly with the patient’s breathing effort. As the patient demands for a new breath Vivo 30 responds immediately and delivers the pressure required.

An intuitive user-interface with Home and Clinical modes enables both the patient and clinician to adjust the desired settings.

An integrated heated humidifier comfortably moistens the air.

An attractive, appealing design and logical user interface make Vivo 30 an integrated part of any home environment.

Settings/Performance
  • PCV (Pressure Control Ventilation)
  • PSV (Pressure Support Ventilation)
  • CPAP (Continuous Positive Airway Pressure)
  • Device Modes: Clinical, Home
  • IPAP/EPAP: 4 to 30 cmH2O / 2 to 20 cmH2O or IPAP -2 cmH2O
  • CPAP: 4 to 20 cmH2O
  • Breath Rate: 4 to 40 BPM
  • Inspiratory Time: 0.3 to 5 s
  • Rise Time: 1 to 9
  • Inspiratory Trigger (eSync): 1 to 9, Off
  • Expiratory Trigger: 1 to 9
  • Ramp function: 10 to 60 min
  • Humidification: 1 to 9, (10 to 30 mgH2O/L)
  • Audible alarm: 1 to 9
  • Maximum flow: > 200 litres/min
  • Minimal Inspiratory Time Off,: 0.3 to 3.0 s
  • Maximal Inspiratory Time: 0.3 to 3.0 s, Off
  • Noise level (at 10 cmH2O constant pressure): Less than 30 dB(A)
Indications
  • Pressure; bar graph and digital data: 0 to 30 cmH2O
  • Calculated tidal volume accuracy: ± 20% or ± 20ml (witchever is the greatest)
  • Leakage calculation accuracy: ± 20%
  • Total Breath Rate: 0 to 60 BPM (Breaths Per Minute)
  • I/E Ratio: 1 :10 to 10 :1
  • Alarms
  • Audible Alarm Level: 35 to 85 dB(A)
  • Low/High pressure alarm, Low volume alarm, Low/High leakage alarm, Low external DC Warning, Low power alarm, Internal function failure alarms.
  • Power Supplies
  • Mains supply: 100 to 240 V AC
  • External battery: 12/24 V DC
  • Operating Conditions
  • Recommended mask leakage: 20 to 40 litres/min at 10 cmH22O
  • Minimum mask leakage: 12 litres/min at 4 cmH2O
  • Dimensions
  • W x H x D: 190 x 225 x 223 mm
  • Weight: 3.1 kg
  • Air outlet: 22 mm male conical standard connector